4-28
Solar 8000M/i patient monitor
2026265-075C
Maintenance and checkout:
Checkout procedure
10. Set the
LIGHT
and
MODULATION
switches as follows:
Set the
LIGHT
switch to
LOW
.
Set the
MODULATION
switch to
HIGH
.
Set the
RATE
switch to
201
.
11. Verify the following SpO2 readings for saturation and pulse rate:
Saturation (%): 81 ± 2
Rate (bpm): 201 ± 3% (195 to 207)
NBP calibration check
1. Connect the acquisition device to the patient monitor.
2. Disconnect all parameter cables.
3. Apply power to the patient monitor.
4. Make sure the power indicator on the acquisition device’s front panel
is ON (green).
WARNING
Do not
put the NBP cuff around a human limb during the
checkout procedure due to the potential for injury. When
the NBP cuff is used in this procedure, it must be tightly
wrapped around a rigid cylinder or pipe.
5. Connect a manometer and NBP cuff to the NBP connector on the
front of the acquisition device.
Connect to
acquisition device
454A
Содержание Solar 8000M/i
Страница 8: ...vi Solar 8000M i patient monitor 2026265 075C ...
Страница 9: ...2026265 075C Solar 8000M i patient monitor 1 1 1 Introduction ...
Страница 16: ...1 8 Solar 8000M i patient monitor 2026265 075C Introduction Service information ...
Страница 17: ...2026265 075C Solar 8000M i patient monitor 2 1 2 Equipment overview ...
Страница 39: ...2026265 075C Solar 8000M i patient monitor 3 1 3 Installation ...
Страница 52: ...3 14 Solar 8000M i patient monitor 2026265 075C Installation Ethernet communication ...
Страница 53: ...2026265 075C Solar 8000M i patient monitor 4 1 4 Maintenance and checkout ...
Страница 92: ...4 40 Solar 8000M i patient monitor 2026265 075C Maintenance and checkout Repair Log ...
Страница 93: ...2026265 075C Solar 8000M i patient monitor 5 1 5 Troubleshooting ...
Страница 125: ...2026265 075C Solar 8000M i patient monitor 6 1 6 Configuration ...
Страница 141: ...2026265 075C Solar 8000M i patient monitor 7 1 7 Field replaceable units ...
Страница 145: ...2026265 075C Solar 8000M i patient monitor 7 5 Field replaceable units Hardware kit ...
Страница 152: ...7 12 Solar 8000M i patient monitor 2026265 075C Field replaceable units Disassembly ...
Страница 153: ...2026265 075C Solar 8000M i patient monitor A 1 A Appendix A Technical specifications ...
Страница 180: ...A 28 Solar 8000M i patient monitor 2026265 075C Appendix A Technical specifications Solar 8000M i display ...
Страница 181: ...2026265 075C Solar 8000M i patient monitor 8000i Patient Monitor B 1 B Appendix B Electromagnetic compatibility ...
Страница 191: ......