Section 14 — Troubleshooting
266
Ethernet Connection Lost in Windows Vista
If you are using Windows Vista, you may lose an Ethernet
communication with your instrument after communicating via USB. This
is the result of network protocol incompatibility. To resolve this problem,
disable support for the IPv6 network protocol in Vista:
1.
Select
Start > Control Panel > View network status
under
Network
and Internet
.
2.
Click
Manage network connections
on the left.
3.
Right-click the
Local Area Connection #X
icon > Select
Properties
.
TIP: You can identify the correct icon by the subheading vbX #X, if
#X is the USB port number.
4.
Deselect the checkbox beside
Internet Protocol Version 6
(TCP/IPv6)
under
This connection uses the following items
.
5.
Click
OK
.
NOTE: Disabling IPv6 will prevent you from using Windows
Meeting Space or any application that relies on the Windows Peer-
to-Peer Networking platform or the Teredo transition technology. If
you wish to use these applications, re-enable IPv6 support.
Ethernet instrument connections will be unaffected by the
reactivation of IPv6 support. This is because the compatibility
problem is resolved by the initial performance of the procedure
above.
Содержание SCOUT100-Ex
Страница 1: ...SCOUT vbSeries Instrument Reference Guide SCOUT100 Ex SCOUT140 Ex vb5 vb6 vb7 vb8 25 September 2014 ...
Страница 18: ...Section 1 Introduction 7 Installation Control Drawing for Hazardous Locations ...
Страница 20: ...Section 1 Introduction 9 Installation Control Drawing GE2214VBX for ATEX Hazardous Locations ...
Страница 21: ...Section 1 Introduction 10 IECEx ATEX Declaration of Conformity ...
Страница 40: ...Section 2 Instrument Fundamentals 29 Viewing hint labels Accessing more detailed help ...
Страница 44: ...Section 2 Instrument Fundamentals 33 Using the advanced options ...
Страница 54: ...Section 3 Sensors 43 Selecting the appropriate voltage for your tachometer and machine ...
Страница 85: ...Section 4 Measurement Types and their Uses 74 Selecting the required parameter settings ...
Страница 111: ...Section 4 Measurement Types and their Uses 100 Using the keypad to enter measurements ...
Страница 116: ...Section 4 Measurement Types and their Uses 105 Example of tracked measurement ...
Страница 120: ...Section 4 Measurement Types and their Uses 109 Saving multi channel measurements ...
Страница 124: ...Section 5 Set Measurement Options 113 Spectral lines submenu ...
Страница 164: ...Section 8 Send and Receive Information 153 Transferring recordings to Ascent ...
Страница 175: ...Section 8 Send and Receive Information 164 Selecting your Wi Fi region Creating a Wi Fi connection ...
Страница 194: ...Section 9 Working with Routes 183 Linear Speed ...
Страница 241: ...Section 11 Balancing Rotors 230 Splitting a weight ...
Страница 247: ...Section 12 Balancing Walkthrough 236 Taking an initial reading ...