Section 4 — Measurement Types and their Uses
96
To execute an actual impact test in
Run
mode:
1.
Exit
Trial
recording mode.
TIP: The recording mode in use displays at the top of the screen
(Trial or Run).
2.
Use your impact hammer to impact the relevant point on the
machine you are measuring.
3.
Review the message that displays on the screen:
•
Press
Save
when you have finished taking all your impact
tests at this location.
NOTE: To help you keep track of where you are in the recording
sequence, your instrument automatically increments the point
names each time you save a recording. It also supports a point
sequence that includes a number and text — E.g. Motor1, Motor2,
etc. However, simply using numbers is quicker and easier than
using an alphanumeric name for each point.
•
Press
if you are satisfied with the impact you have
executed and wish to perform another impact.
TIP: Alternatively you can simply perform another trial impact —
The instrument is ready and waiting for you.
•
Press
to discard the impact you have just taken.
NOTE: You can only discard the most recent impact taken. It is not
possible to discard older impacts.
NOTE: You can press
to toggle between displaying impact and
response waveforms; impact and response spectra; and
Coherence/FRF/Phase graphs. All these graphs display in real time
while you are acquiring data.
TIP: Press
to access more detail about the warning message
while it displays.
TIP: You do NOT need to press a button to retain an impact. You
can simply execute a new impact to achieve this.
Содержание SCOUT100-Ex
Страница 1: ...SCOUT vbSeries Instrument Reference Guide SCOUT100 Ex SCOUT140 Ex vb5 vb6 vb7 vb8 25 September 2014 ...
Страница 18: ...Section 1 Introduction 7 Installation Control Drawing for Hazardous Locations ...
Страница 20: ...Section 1 Introduction 9 Installation Control Drawing GE2214VBX for ATEX Hazardous Locations ...
Страница 21: ...Section 1 Introduction 10 IECEx ATEX Declaration of Conformity ...
Страница 40: ...Section 2 Instrument Fundamentals 29 Viewing hint labels Accessing more detailed help ...
Страница 44: ...Section 2 Instrument Fundamentals 33 Using the advanced options ...
Страница 54: ...Section 3 Sensors 43 Selecting the appropriate voltage for your tachometer and machine ...
Страница 85: ...Section 4 Measurement Types and their Uses 74 Selecting the required parameter settings ...
Страница 111: ...Section 4 Measurement Types and their Uses 100 Using the keypad to enter measurements ...
Страница 116: ...Section 4 Measurement Types and their Uses 105 Example of tracked measurement ...
Страница 120: ...Section 4 Measurement Types and their Uses 109 Saving multi channel measurements ...
Страница 124: ...Section 5 Set Measurement Options 113 Spectral lines submenu ...
Страница 164: ...Section 8 Send and Receive Information 153 Transferring recordings to Ascent ...
Страница 175: ...Section 8 Send and Receive Information 164 Selecting your Wi Fi region Creating a Wi Fi connection ...
Страница 194: ...Section 9 Working with Routes 183 Linear Speed ...
Страница 241: ...Section 11 Balancing Rotors 230 Splitting a weight ...
Страница 247: ...Section 12 Balancing Walkthrough 236 Taking an initial reading ...