Section 1 — Introduction
8
IECEx / ATEX Certification — SCOUT Instruments
SCOUT100-Ex and SCOUT140-Ex portable vibration analyzers are
IECEx /
ATEX certified
— Identified by the
logo on the rear instrument
label.
If you need to use your instrument in hazardous areas, please ensure it
is the appropriate variant for your country. Refer to the following pages
for the hazardous area usage instructions.
The
logo on the rear label indicates your instrument has been
approved for use in hazardous locations with the following classification:
II 3 G Ex ic IIB T4 Gc
•
Group
II
= non-mining
•
Category
3
= Zone 2
•
Atmosphere
G
= Gas
•
Protection method
ic
= intrinsic safety for zone 2
•
Explosion Group
IIB
= gases such as Ethylene
•
Temperature Class
T4
= surface will not exceed 135°C
•
Protection level
c
= Zone 2
The approval covers use in Europe (under ATEX) and any other countries
that recognize the IECEx certification.
When using a certified instrument in a hazardous location, the
installation must comply with the following Installation Control Drawing
for ATEX Hazardous Locations GE2214VBX , including all its footnotes
and warnings.
Your instrument will be supplied, complete with appropriate sensors for
use in Zone 2 environments. As indicated in the following drawing, you
can use alternative sensors, provided these are certified for use in these
locations, and provided their entity parameters meet the limits shown in
the drawing.
Содержание SCOUT100-Ex
Страница 1: ...SCOUT vbSeries Instrument Reference Guide SCOUT100 Ex SCOUT140 Ex vb5 vb6 vb7 vb8 25 September 2014 ...
Страница 18: ...Section 1 Introduction 7 Installation Control Drawing for Hazardous Locations ...
Страница 20: ...Section 1 Introduction 9 Installation Control Drawing GE2214VBX for ATEX Hazardous Locations ...
Страница 21: ...Section 1 Introduction 10 IECEx ATEX Declaration of Conformity ...
Страница 40: ...Section 2 Instrument Fundamentals 29 Viewing hint labels Accessing more detailed help ...
Страница 44: ...Section 2 Instrument Fundamentals 33 Using the advanced options ...
Страница 54: ...Section 3 Sensors 43 Selecting the appropriate voltage for your tachometer and machine ...
Страница 85: ...Section 4 Measurement Types and their Uses 74 Selecting the required parameter settings ...
Страница 111: ...Section 4 Measurement Types and their Uses 100 Using the keypad to enter measurements ...
Страница 116: ...Section 4 Measurement Types and their Uses 105 Example of tracked measurement ...
Страница 120: ...Section 4 Measurement Types and their Uses 109 Saving multi channel measurements ...
Страница 124: ...Section 5 Set Measurement Options 113 Spectral lines submenu ...
Страница 164: ...Section 8 Send and Receive Information 153 Transferring recordings to Ascent ...
Страница 175: ...Section 8 Send and Receive Information 164 Selecting your Wi Fi region Creating a Wi Fi connection ...
Страница 194: ...Section 9 Working with Routes 183 Linear Speed ...
Страница 241: ...Section 11 Balancing Rotors 230 Splitting a weight ...
Страница 247: ...Section 12 Balancing Walkthrough 236 Taking an initial reading ...