Section 5 — Set Measurement Options
121
Equivalent Fmax
You can use equivalent Fmax to control the waveform sampling rate by
expressing it in terms of the maximum frequency of interest. For
example, if you want a waveform with frequency content (Equivalent
Fmax) up to 1 kHz, the instrument will automatically sample at 2.56 kHz.
Changing the Equivalent Fmax directly affects the sampling rate, which
has an inverse effect on the duration of the waveform.
Number of Samples
The resolution of the waveform increases with the number of samples
used. The more samples, the more information the waveform contains.
However, more samples in a waveform means more memory is used to
store the waveform.
TIP: You can let your instrument set the number of samples and
duration for you. To access this feature, simply set the Equivalent
Fmax so that it covers your required frequency range. Your
instrument will select the maximum number of samples and
duration that can be used with this Fmax. You can then adjust
these values as required.
Duration
The duration of a waveform is its recording time. The duration value
depends on the number of samples selected. Increasing the number of
samples increases the duration.
Additional Measurement Parameters
The following additional measurement parameters are available:
•
Bandwidth
— Available with Demod Spectrum. The bandwidth
you select determines which range of frequencies the
demodulation process will use.
•
Machine speed
— Available with Cross Channel Phase. When
you measure cross channel phase, ensure you specify the
frequency you are investigating. Typically, this is the machine
running speed.
Содержание SCOUT100-Ex
Страница 1: ...SCOUT vbSeries Instrument Reference Guide SCOUT100 Ex SCOUT140 Ex vb5 vb6 vb7 vb8 25 September 2014 ...
Страница 18: ...Section 1 Introduction 7 Installation Control Drawing for Hazardous Locations ...
Страница 20: ...Section 1 Introduction 9 Installation Control Drawing GE2214VBX for ATEX Hazardous Locations ...
Страница 21: ...Section 1 Introduction 10 IECEx ATEX Declaration of Conformity ...
Страница 40: ...Section 2 Instrument Fundamentals 29 Viewing hint labels Accessing more detailed help ...
Страница 44: ...Section 2 Instrument Fundamentals 33 Using the advanced options ...
Страница 54: ...Section 3 Sensors 43 Selecting the appropriate voltage for your tachometer and machine ...
Страница 85: ...Section 4 Measurement Types and their Uses 74 Selecting the required parameter settings ...
Страница 111: ...Section 4 Measurement Types and their Uses 100 Using the keypad to enter measurements ...
Страница 116: ...Section 4 Measurement Types and their Uses 105 Example of tracked measurement ...
Страница 120: ...Section 4 Measurement Types and their Uses 109 Saving multi channel measurements ...
Страница 124: ...Section 5 Set Measurement Options 113 Spectral lines submenu ...
Страница 164: ...Section 8 Send and Receive Information 153 Transferring recordings to Ascent ...
Страница 175: ...Section 8 Send and Receive Information 164 Selecting your Wi Fi region Creating a Wi Fi connection ...
Страница 194: ...Section 9 Working with Routes 183 Linear Speed ...
Страница 241: ...Section 11 Balancing Rotors 230 Splitting a weight ...
Страница 247: ...Section 12 Balancing Walkthrough 236 Taking an initial reading ...