Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5
pim 2337741–100
40
1-4
Radiation Protection
Because x–ray equipment produces radiation, special precautions may be needed or special site
modifications may be required. The General Electric Company does not make recommendations
regarding radiation protection. It is the purchasers’ responsibility to consult a radiation physicist for
advise on radiation protection in x–ray rooms.
1-5
Mobile requirements
To prevent any potential susceptibility and system perturbation, the Mobile Innova trailer should not
be located and operated near the following:
High voltage transformers,
Antennas (radio, TV or telecom)
Other system behind the hospital wall that could disturb the Mobile Innova (X–Ray, CT, MR, ...).
Other system behind the hospital wall that could be disturbed by the Mobile Innova (nursery,
emergency room, ...)
Any other equipment, which may disturb the Mobile Innova.
1-6
Audible noise
Less than 62 dB (A) at 1 meter for a Innova LC Positioner
Limited to 50 dB (A) at 1 meter for Omega 4 table.
Limited to 65 dB (A) at 1 meter for VMP Cabinet.
Limited to 57 dB (A) with a “low noise” cover or to 65 dB (A) with a “high ambient” cover at 1 m
for the COOLIX 2200.
Limited to 65 dB (A) at 1 meter for ACAB Cabinet
Limited to 60 dB (A) at 1 meter for Digital Detector Chiller
Less than 52 dB (A) at 1 meter for a DL Flat panel.
Limited to 60dB(A) at 1 meter for optional UPS.
1-7
Windows and curtains
When the examination room has a window with an aperture outside of the controlled light area (day
light, other...) a curtain has to maintain the light intensity under a limit fixed to 150 lux.