CHAPTER 1 – INTRODUCTION
1
OBJECTIVE AND SCOPE OF THIS MANUAL
This document is intended as a guide and information resource to properly plan and prepare a site for
the installation of an Innova 2000 system.
In addition, this document provides references to the pre–installation documents of the various
product included with an Innova 2000
System.
These documents are intended to assist the Installation Specialist and the Site Planner in properly
preparing a site for the installation of this system.
It provides preinstallation data, such as site preparation prior to the delivery of the Innova 2000
System, environmental and electrical requirements and some additional planning aids.
MAKE SURE THE ROOM PREPARATION COMPLIES WITH LOCAL
REGULATIONS AS THE PIM IS NOT INTENDED TO REFLECT ALL OF THEM.
2
DESCRIPTION OF INNOVA 2000 SYSTEM
An Innova 2000
system is used for cardiography cases with several options.
An Innova 2000 system is first offered in one configuration.
2-1
Description of the Innova LC positioner
A complete Innova LC positioner comprises only three parts.
Innova 2000 positioner/titan cabinet,
Innova 2000 positioner including an X–ray tube housing with a SIEMENS collimator and a 22 cm
Revolution Digital Detector.
Multiaxis control (smart /or Smart Box).
2-2
Description of the Innova Patient table
Innova 2000 Systems can be supplied with an OMEGA IV Compact patient table which includes a
table side control with Contour Filter or an OMEGA V Long table.
Innova patient tables are offered following the needs of customer. The main
recommendation is the location from Innova LC positioner. Omega IV
Compact or Omega V Long tables are basically located to LC at 1395 mm
but for long examination, Omega V can be located at 1278 mm. Have to be
defined before site preparation.