Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5
pim 2337741–100
33
1-2
Equipment Heat Output (Dissipation)
Refer to Table 15. To obtain heat output information for components not specified inTable 15, refer to
the appropriate component Pre–Installation Manual listed in Chapter 2.
Table 15 – INNOVA 2000 SYSTEM HEAT OUTPUTS (BY COMPONENT)
PRODUCT OR COM-
HEAT OUTPUT (BTU/HR)
HEAT OUTPUT (WATTS)
PRODUCT OR COM
PONENT
STANDBY
IN–USE
STANDBY
IN–USE
Innova LC Positioner
negligible
1166
negligible
342
POSTIT
Cabinet
5600
11123
1640
3260
Omega
Patient table
negligible
negligible
Maxiray 150 V5
X–Ray tube assy
negligible
1,000
negligible
300
Innova Digital Detector
ACAB
Cabinet
2500/2800W
VMP
Cabinets
3400
5100
1000
1500
Tube Chiller
5120
20470(1/2)
1500
6000W
Digital Detector Chiller
250W
Innova 2000
Operator console
181
194
53
57
DL user area
1000
300
Note 1: Air intake requirements 1360 cubic meter per hour
2: For more details, consult appropriate pre–installation manual