Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5
pim 2337741–100
63
Illustration 24 – X–RAY CHILLER DIMENSIONS
The maximum elevation difference between X–Ray Tube Housing and
Chiller is around 3 meters (10 ft).
Exact dimensions of the chiller depends on the model type.
Required floor space depends on ambient room temperatures. When in doubt, allow for maximum
floor space.
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏÏÏ
140 cm MINIMUM
(55 in)
165 cm MINIMUM
(65 in)
95 cm MINIMUM
(37 in)
105 cm MINIMUM
(41 in)
72 cm MAXIMUM
(28 in)
83 cm MAXIMUM
(33 in)
13 cm MINIMUM
(5 in)
FRONT
Floor space needed with room ambient
temperatures greater than 25C
Floor space needed with room ambient
temperatures of 25C or less
CAUTION