Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5
pim 2337741–100
136
JOB CARD PIST044–007 – System facility power
2
2 of 8
5
SCOPE
Invasive procedure room shall have all exposed metal parts
that are likely to become energized,
grounded to an approved grounding bus called room ground point. The room ground point is located:
in the distribution panel (PDB) for Innova 2000 system and for any other medical devices powered
in the examination room (refer to Illustration 2).
in the Atlas Cabinet for Innova 2000 subsystems.
All room outlets and emergency power sources in the room shall isolated ground receptacles with the
primary grounding coming from the power source and a secondary ground bonded to the room
ground point. For the receptacle or electrical box which powers the injectors power module there must
be one ground wire back to the room ground point even if the power module is in a separate room.
All ground wire impedances shall be less than 0.1
when measured to the room ground point.