Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5
pim 2337741–100
48
Illustration 16 – CABLE CURVATURE
(FLOOR)
LCA BASEPLATE
ISOCENTER
WATER LINES
THRU–FLOOR FITTING
(FLOOR)
TABLE SIDE
Note:
In case of thru–floor cabling, if the electrical cables are coming from the head side, they will
need to have a minimum curvature with a minimum radius of 150 mm (6”).
In any other cases (i.e. flush floor) no such curvature is allowed.
Mini radius = 150 mm (6 inches)
Note:
In case of on grade cabling, because of the minimum curvature constraint of 150 mm (6”),
the cable will have to come from the side between Innova LC Positioner and patient table.
Electrical cables coming
from positioner cabinet
THRU–FLOOR
LCA Baseplate
Cable conduit
(flush floor
interface fitting)
4 water pipes between
Chillers
and LCA H1C positioner
TABLE
SIDE
Electrical cables
coming from cabinets
ON
GRADE