Aisys
2-36
04/08 M1046983
2.11 Gas flow through the anesthesia machine
2.11.1 Overview
Refer to Figure 2-27 and Figure 2-28.
Gas supplies
Gas comes into the system through a pipeline (
1
) or cylinder (
6
) connection. All
connections have indexed fittings, filters, and check valves (one-way valves). Pressure
transducers monitor the pipeline (
2
) and cylinder (
7
) pressures.
The O
2
supply failure alarm is derived from the O
2
pipeline and the O
2
cylinder pressure
transducer inputs.
A primary regulator (
8
) decreases the cylinder pressures to approximately pipeline
levels. A pressure relief valve (
3
) helps protect the system from high pressures.
To help prevent problems with the gas supplies:
• Install yoke plugs on all empty cylinder connections.
• When a pipeline supply is adequate, keep the cylinder valve closed.
Gas flow
Pipeline or regulated cylinder pressure supplies O
2
or Air directly to the ventilator
engine (
4a
or
4b
) and as pilot pressure (
4
) for the SCGO
assembly (
E
). Connection points are also available for venturi
suction (
5a
or
5b
) drive gas supply. An additional O
2
regulator (
18
) decreases the
pressure for the O
2
Flush valve (
19
) and the auxiliary O
2
flowmeter (
24
).
The O
2
Flush valve supplies high flows of O
2
to the fresh gas outlet (
22
or
23
) through
the SCGO/ACGO assembly (
E/F
). The flush pressure switch (
20
) monitors activation of
the flush valve.
Gas mixing
Under normal conditions, with the system switch (
10
) in the On position, the Alternate
O
2
Disable valve (
13
) is energized to block alternate O2 flow. Normal gas flows are
enabled through their respective selector valves (
11
). The system controls gas flow
through the flow control valves (
12
) and derives the individual flow rates through the
hot-wire anemometers (
14
).
Under system failure conditions (or if Alt O
2
is selected), the normally-open Alternate
O
2
Disable valve (
13
) allows delivery of O
2
through the Alternate O
2
Flowmeter when
the system switch is in the On position.
Mixed gas
The mixed gas (
15
) flows through the electronic vaporizer (
D
) to the
SCGO/ACGO assembly (
E/F
). A pressure relief valve (
17
) on the electronic vaporizer
limits the maximum outlet pressure.
The SCGO assembly (
E
) directs the mixed gas to the selected circuit:
22
(ABS-circle) or
23
(to Inspiratory port of ABS). On SCGO assemblies, a relief valve
(
21
) limits pressure in the breathing system to approximately
150 cmH
2
O.
The ACGO assembly (
F
) directs the mixed gas to the selected circuit:
22
(ABS-circle) or
23
(external ACGO port).
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......