7 Troubleshooting
M1046983 04/08
7-69
PCB Alarm On:
Monitor Current
Active w/AC Mains
Indicates the Battery
backup current to monitor is
active while AC supply is OK.
PCB
Check AC power connection to anesthesia monitor and circuit breaker.
If problem continues, replace anesthesia monitor.
PCB Saved Alarm Off:
1 min time left
A message or failure
displayed during the system
checkout.
PCB
Leave the system plugged into AC Mains for 24 hours. If issue persists, replace batteries.
PCB Saved Alarm Off:
AC Mains failure
Indicates AC Power
Removed.
PCB
No Service Action Required.
PCB Saved Alarm Off:
Blk chrg time exceed
in stdby
Indicates the batteries were
being Bulk Charged for a
period greater than 12
hours while the system was
in the Standby state.
PCB
Replace Batteries.
Replace the Power Controller Board.
PCB Saved Alarm Off:
DC-DC power module
failure
Indicates greater than 0.25
amps current draw out of
the batteries for 2 minutes
while AC Mains is connected
to the machine.
PCB
Replace Batteries.
Replace the Power Controller Board.
PCB Saved Alarm On:
1 min time left
A message or failure
displayed during the system
checkout.
PCB
Leave the system plugged into AC Mains for 24 hours. If issue persists, replace batteries.
PCB Saved Alarm On:
AC Mains failure
Indicates AC Power
Removed.
PCB
No Service Action Required.
PCB Saved Alarm On:
DC-DC power module
failure
Indicates greater than 0.25
amps current draw out of
the batteries for 2 minutes
while AC Mains is connected
to the machine.
PCB
Replace Batteries.
Replace the Power Controller Board.
Error Log Entry
Alarm Text
Condition (Basic info)
Priority Source Enabling Criteria
Action/Troubleshooting
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......