2 Theory of Operation
M1046983 04/08
2-19
Power and
Communication to
other Subsystems
The Anesthesia Controller Board provides power and communication to:
Ventilator Interface Board (VIB):
• Power = 12.5VDC
• Communication = RS422
Gas Mixer:
• Power = 12.5VDC
• Communication = RS422
Agent Delivery Board (ADB):
• Power = 12.5VDC
• Communication = RS422
M-Gas Power Board:
• Power = 12.5VDC
• Note: Communication to and from the M-Gas is via the Display Unit
Gas Pressure
Transducers:
Gas Pressure Transducers monitor the Pipeline pressures of all gases connected to the
system. Gasses measured can include O
2
, Air, and N
2
O. Their measurement range is
form 0 to 697 kPa (101 psi). Though these transducers are located near the machine
pipeline connection points, they are maintained and monitored by the Anesthesia
Controller Board.
Pressure transducers also monitor all cylinder pressures. Gasses measured can
include O
2
, second O
2
,
Air, and N
2
O. Their range is:
• from 0 to 27580 kPA (4000 psi) for O
2
and Air
• from 0 to 9805 kPa (1422 psi) for N2O
The transducers are located on each cylinder yoke; they are maintained and monitored
by the Anesthesia Controller Board.
Alt O
2
Switch
The Anesthesia Controller Board monitors the position of the Alt O2 Switch.
Display Unit
The Display Unit receives power from and has limited communication with the PCB. This
limited communication allows for data like power supply status and voltage readings to
be displayed in service mode.
The MAJOR communication responsible for getting clinical settings to the system and
the ability to display parameters occurs between the Anesthesia Controller Board and
Display Unit. The Anesthesia Controller Board and Display Unit communicate via
RS422.
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......