Aisys
7-52
04/08 M1046983
CASSETTE
TEMPERATURE
EEPROM FAILURE
Try another cassette.
Schedule service.
Cassette temperature sensor
calibration data EEPROM
read error or cassette
temperature sensor hardware
revision data EEPROM read
error or software compatibility
failure.
Medium
De-
escalating
AC
Check connection of Cassette Temperature Sensor to the ADB.
Cycle power.
If the problem persists replace and retest, in the following order (Also see Section 7.10.):
• Cassette temp sensor
• ADB
CAL DATA FAILURE IN
EEPROM
Service calibration
advised
Default cal data is being used
due to corrupt data in cal
region.
Low
AC
Perform complete service level calibrations (ventilator).
CHECK FLOW SENSOR
Check flow sensors
During Mechanical breaths,
the measured flow for 6
consecutive breaths, to and
from the patient, does not
meet certain criteria.
No or negative flow on Insp
flow sensor during inspiration
or negative flow on Exp flow
sensor.
Medium
AC,
Vent
In-range flow data
available during
mechanical ventilation
Check flow sensor connections.
Check the breathing circuit.
Check VIB sensor tubing for leaks.
Perform flow sensor calibration.
Check Insp/Exp check valves.
Check/Replace flow sensors.
CLOSED LOOP
CONTROL FAILURE
Check cassette. Set
agent.
Closed loop controller was
unable to control output flow
from the cassette.
Medium
De-
escalating
AC
System state is
Therapy
Run PC Service App Vaporizer Test (Section 12.10.2) to verify hardware performance.
Replace and retest in the following order (Also see Section 7.10.):
• Proportional Valve
• ADB
This error can also be caused by the following:
• Liquid Flow Prevention Valve actuated. Look for occurrences of CONDENSATION CONDITIONS EXIST
and CASSETTE OVERFILL DETECTED errors preceding this error.
• Cassette held in place, but not latched, during agent delivery. It is possible to hold the cassette in a
position such that the cassette ID is recognized but the cassette valves are not open, preventing flow.
• Check cassette travel to ensure deformation or under sizing of cassette bay is not causing undue
restriction.
Error Log Entry
Alarm Text
Condition (Basic info)
Priority Source Enabling Criteria
Action/Troubleshooting
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......