3 Checkout Procedure
M1046983 04/08
3-3
3.2.2 Machine Check
The machine check runs automatically and beeps to indicate when it is
finished or if interaction is required.
The
Machine Check
does a:
•
Machine Check - System
check (Ventilator Circuit),
•
Machine Check- Circuit
check (Bag Circuit),
• and a
Machine Check - Circuit O2
cell check (if circuit O
2
cell is present).
When one of the checks is completed, the system will transition to the next
check.
1. Turn the System switch to On.
2. Select
Machine Check
and follow the instructions.
3. If a check fails, follow the instructions to perform a recheck or accept the
results.
3.2.3 Machine Check -
System (Ventilator
Circuit Testing)
The
Machine Check-System
checks the Bag/Vent switch, proper gas supply
pressures, ventilator operation and leak, battery and electrical power, circuit
compliance, flow control operation, and vaporizer operation. This is a two-
step check.
Note
The
Machine Check-System
can be run with a test cassette to invoke
extended diagnostics (Agent Delivery test will take several minutes longer).
1. Set the Bag/Vent switch to Vent.
2. Open the patient Y.
3. (ACGO option only.) Set the ACGO switch to Circle.
4. Select
Start
. The display shows the checks being run.
•
The system beeps when this portion of the check is done.
•
The results are shown on the display.
5. Make sure the bellows is fully collapsed.
6. Occlude the patient Y.
7. Select
Continue
. The display shows the checks being run.
8. When the check passes, the system will transition to the next step.
3.2.4 Machine Check -
Circuit (Bag Circuit
Testing)
The
Machine Check-Circuit
checks the Bag/Vent switch, proper gas supply
pressures, airway pressure measurement transducer, APL valve, and manual
circuit leak.
1. Occlude the patient Y.
2. Set Bag/Vent switch to Bag.
3. Set the APL valve halfway between 30 and 70.
4. (ACGO option only.) Set the ACGO switch to Circle.
5. Select
Start
. The display shows the checks being run.
•
The system beeps when the check is done.
•
The results are shown on the display.
6. When the check passes, the system will transition to the next step.
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......