Aisys
7-60
04/08 M1046983
MAN CASS OVER
UNDER TEMP
Check cassette. Set
agent.
Either manifold temperature
reading or cassette
temperature reading outside
of limit.
Medium
De-
escalating
AC
Operating temperature as measured by one of the Electronic Vaporizer temperature sensors was out of
allowed operating range.
MANIFOLD PAW
SENSOR FAIL
Vent Fail. Monitoring
Only
Calibration failure at bootup. Medium
AC,
Vent
In the Service Software / Vent Flow and Pressure Diagnostics, verify the Manifold Flow counts is
800 ± 250.
Disconnect the White in-line connector in the Manifold Pressure. If the counts return within specified
range, check for occlusions in the Bulkhead harness.
If the counts do not return within the specified range, replace the VIB.
MANIFOLD
PRESSURE SENSR
FAILURE
Indicates a calibration
failure at bootup.
ACB
In the Service Software / Vent Flow & Pressure Diagnostics, verify the Manifold Flow counts is
800 ± 250.
Disconnect the White in-line connector in the Manifold Pressure. If the counts return within specified
range, check for occlusions in the Bulkhead harness.
If the counts do not return within the specified range, replace the VIB.
MANIFOLD SENSOR
CAL ERROR
Indicates the Manifold
Pressure zero failed.
ACB
Could be caused by bad span calibration or leaky vent Insp Flow Valve.
MGAS CHECK SAMPLE
GAS OUT >20 SEC
Check sample gas out
MGAS SPEC. Continuous
Occlusion Bit set.
Medium
MGAS
MGAS present and
MGAS communicates
continuous occlusion
for 20 seconds
Replace sample line. See AM TRM for further Troubleshooting.
MGAS INLET FILTER
RESIDUE >40 SEC
Replace D-Fend
MGAS SPEC (Residue build-
up on the water trap
membrane. This decreases air
flow).
Medium
MGAS
MGAS present and
MGAS communicates
this the Replace Trap
alarm bit for 40
seconds
Replace D-Fend. See AM TRM for further Troubleshooting.
MGAS LINE BLOCKED
>20 SEC
Sample line blocked
MGAS SPEC states The
sample tubing inside or
outside the monitor blocked,
or the water trap is occluded.
Medium
MGAS
MGAS present and
MGAS communicates
this the continuous
occlusion alarm for 20
seconds
Replace sample line. See AM TRM for further Troubleshooting.
Error Log Entry
Alarm Text
Condition (Basic info)
Priority Source Enabling Criteria
Action/Troubleshooting
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......