B30 Patient Monitor
8-16
Document no
.
2044677-001
Recorder
17. Paper compartment cleaning
Open the paper compartment hatch and take out the paper roll, if installed.
Remove any paper chaff from the compartment.
Clean the thermal printhead and the small glass window in front of the static brush with a
cotton swab dipped in isopropyl alcohol. Avoid contact with the rubber paper roller.
NOTE: Be careful to limit the application of alcohol to the thermal printhead and the window.
Leave the paper compartment empty and close the hatch.
"
Configure the monitor screen so that all required parameters are shown, for example:
Monitor Setup
- Screen 1 Setup - Waveform Fields - Field 1 - ECG1
Field 2 - ECG2
Field 3 - P1
Field 4 - P2
Field 5 - PLETH
Field 6 - OFF
Plug in the Hemodynamic Module. Connect a patient simulator to the module and check
that all connected parameters are shown on the screen.
Preset recording settings:
Record/Print
-
Record Waveforms
-
Waveform 1 - ECG1
Waveform 2 - P1
Waveform 3 - P2
Delay - Off
Paper Speed - 6.25 Mm/S
Length --> 30 S
Record Trends - Graphic Trend 1 - Hr
Graphic Trend 2 - P1
Monitor Setup
-
Install/Service
(password 16-4-34)
- Installation - Printer & Recorder -
Default Trend - Graph.
Display Trends - Time Scale - 2 h
18. Paper recognition
Press the
Start/Stop
key on the monitor side panel. Check that the message “Recorder:
Out of paper’ appears on the screen.
"
19. Cover state recognition
Open the paper compartment cover. Check that the previous message changes to
‘Recorder: Cover open’.
Install a paper roll and close the cover. Check that the message ‘Recorder: Cover open’
disappears from the screen.
"
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......