B30 Patient Monitor
4-12
Document no
.
2044677-001
‘Cuff occlusion’ message
1. Cuff and/or hose occluded.
Reason:
- cuff tube kinked
- Straighten tube.
- tubes inside the shield kinked
- Straighten tubes.
- tubes inside module kinked
- Straighten tubes.
- occlusion inside/outside module - Remove occlusion.
2. Cuff, hose, and tubes OK.
Reason:
- fault in pressure transducer
- Replace the NIBP board.
- fault in A/D converter
- Replace the NIBP board.
- faulty calibration
- Check calibration.
‘Calibration switch on’
message
EEPROM protection has been
handled by pressing module buttons
Auto ON/OFF
and
Start/Cancel
simultaneously for 3 seconds.
Enables setting the protection OFF in the
Calibration
menu. Press the buttons
again if you are not going to calibrate.
‘Calibration not protected’
message.
Calibration protection is set to OFF.
Set the protection ON in the NIBP
Calibration
menu.
Problem
Cause
What to do
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......