B30 Patient Monitor
5-4
Document no
.
2044677-001
2.2 Network
2.2.1 Network
Network:
For setting the monitor's network mode:
S5/Unity
NOTE: Only in Chinese language, Unity mode is
available.
2.2.2 Dri Config
The DRI Level:
For setting the monitor's network
communication. The network communication needs
match to the iCentral’s configuration
Transfer Mode
: For choosing the transfer mode:
DRI/ETH or DRI/UDP
Default SLP IP:
For R&D purpose only. (DRI/ETH is not
in use)
Forced Central IP:
For setting IP address, which
according to the iCentral’s configuration. (DRI/ETH is
not in use, it’s can capture the IP automatically)
Forced Central Port:
For setting port address, which
according to the iCentral’s configuration. (DRI/ETH is
not in use, it’s can capture the port automatically)
Debug server IP:
For R&D purpose only.
Software download:
For setting software download
capability.
Save Changes:
Select "Save Changes" to take in use
changes made in this page.
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......