Parameters
5-23
Document no. 2044677-001
4.4.2 NIBP Safety Valve
Test:
Start test
is for starting and
Stop test
for
stopping the Safety Valve test.
Safety Valve Data:
For information on general items
Pressure
,
Zero
,
Protect handle
,
Calibr. prot.
,
+15 V power
,
AD0
to
AD7
as well as
Timeouts
etc., see service data
descriptions in section
.
Max. press
and
2 s after stop
show the measured
values at Safety Valve test.
Safety Valve Test Adult/Infant
Disconnect the pressure manometer from the NIBP
module cuff connector. Connect the NIBP hose and
cuff to the NIBP module cuff connector. Perform the
check with a standard adult cuff that is connected
around some round object, for example a calibration
gas bottle.
Select
Start Test
. Start the adult safety valve test by
pressing the ComWheel. Wait until the pump stops
and the pressure is deflated.
Open the cuff connector or disconnect and connect the cuff connector from the module.
Check the pressure values ‘Max press’ and ‘2 s after stop’ for both transducers. All the values should be
within 270 - 330 mmHg.
Select
ADULT
. Press the ComWheel and check that the text changes now to
INFANT
.
Select
Start Test
and wait until the pump stops and the pressure values on the screen have been updated.
Open the cuff connector or disconnect and connect the cuff connector from the module.
Check that the values ‘Max press’ and ‘2 s after stop’ are all now within 135 to 165 mmHg.
Return to the normal monitoring mode by pressing
Normal Screen
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......