Frame troubleshooting
4-7
Document no. 2044677-001
‘Network board error’
message shows on the
monitor screen
NET section on the CPU board
NET section memory on the
CPU board
The NET section is defective. The board cannot
be used. See network service page for details.
The SRAM memory of the NET section is
defective or uninitialized. The board cannot be
used. See network service page for details.
Other Site View shows no
waveforms
No waveforms are set up for
Monitor-to-Monitor
communication
Run Network Setup to verify current
Monitor-to-Monitor communication setup.
Network printing fails
Print server is busy
Print queue is full
Printer is off-line
Network manager's print server is busy at the
moment and cannot take more print jobs. Try
again after 15 seconds.
There are too many unprinted documents
waiting in the print queue. Check the printer, as
it is not operating properly.
Printer cable is loose, printer is out of paper,
there is a paper jam or the printer is simply
switched to off-line state.
Problem
Cause
What to do
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......