![background image](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170090.webp)
B30 Patient Monitor
3-20
Document no
.
2044677-001
Monitor Setup
Display Brightness
Push the ComWheel and check that the display brightness follows the selected
brightness.
"
4.2.3 Keyboard(s)
1.
Tests with the Command Board:
−
Press the
Monitor Setup
key. Turn the ComWheel in both directions and check that
the cursor in the menu moves correspondingly. Select
Normal Screen
and check
that the menu disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
2.
Check that the clock on the screen shows correct time.
Readjust the time and date, if necessary.
3.
Enter the
Service Log
service menu.
Check the content of the Service Log for possible problems.
"
4.2.4 E-PSM(P)W Module
ECG measurement
1.
Module software (serial numbers)
Enter the service menu:
Monitor Setup
-
Install/Service
(password 16-4-34) -
Service
(password 26-23-8)
Take down the information regarding the module software by selecting
Scroll Vers
and
turning the ComWheel.
"
2.
Communication and memories
Enter the
Parameters -
ECG
service menu.
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check also that the ECG/RESP board memories have passed
the internal memory test, i.e. the RAM, ROM and EEPROM state all OK.
"
3.
Power frequency
Check that the power frequency value is set according to the current mains power
frequency. Change the setting by selecting
Power Freq
, if necessary.
"
4.
Cable recognition
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......