B30 Patient Monitor
5-10
Document no
.
2044677-001
Smart Batt2
If Smart Battx is selected for a battery
NOT connected to the SMBus, the menu
contains the following:
Battx information from Memory and PMC
(This information comes from the DC/DC
Board memory or is measured by the
DC/DC Board. SMBus data in this menu is
not real time, because this battery is not
connected to the SMBus).
NOTE: This page may not contain
information if SMBus has been connected
only to the other battery. Page can be
updated by battery disconnection and
reconnection, if desired.
Rel. State Charge:
Relative state of charge
of the battery. Expressed as a percentage
of Rem. Cap. (mAh) / Full Cap.(mAh).
Remaining Capacity:
Remaining battery
capacity (mAh).
Full Charge Capacity:
Capacity of the
battery when it is fully charged.
Design Capacity:
Theoretical capacity of a
new battery.
Design Voltage
: Theoretical value for nominal voltage of a new battery.
Manufacturer Date (DD:MM:YY):
The date the battery pack was manufactured
Manufacturer Name
: Acronym of the battery pack manufacturer name.
Device Name:
Battery pack model name.
Device Chemistry:
Battery chemistry of the cells used.
Batt1 Voltage:
Battery A voltage measured at the DC/DC Board.
Batt2 Voltage:
Battery B voltage measured at the DC/DC Board.
Batt Current:
Battery current (discharge or charge) for the battery connected to SMBus.
Measured at the DC/DC Board.
Batt Not CHG Temperature:
This is subject to change.
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......