B30 Patient Monitor
8-12
Document no
.
2044677-001
The instructions include a check form (
) which should be filled in when performing
the procedures.
The symbol
"
in the instructions means that the check form should be signed after
performing the procedure.
4.2.1 Recommended tools
NOTE: *)Ensure that the calibrating gas and regulator are functioning properly before
calibration. Perform annual maintenance on the regulator as required. For more information
see
“Calibration gas regulator flow check”
in the
chapter.
4.2.2 Recommended parts
Detach the module box cover by removing the five screws from the locking side of the
module. Be careful with the flex cable, when opening the cover. Remove the grey plastic
shield over the miniC unit.
4.2.3 Inspection
General
1.
Check internal parts
•
all screws are tightened properly
•
all cables are connected properly
•
tubes are not pinched and there are no sharp bends on them
•
all tubes are connected properly
•
EMC covers are attached properly
•
there are no loose objects inside the module
N-F(C)REC:
Open the paper compartment hatch and take out the paper roll, if installed.
Tool
Order No.
Notes
Screwdriver
Pozi-drive, Torx T10
Ambient pressure manometer
Amb. Press. can be checked
from local meteorological
station
Flowmeter
TSI model 4140 recommended
Flow cassette 50/1.1
873812
Calibration gas and the regulator 755580* (gas)
755534
Contains 5 CO
2
and air
E-PSMW / E-PSMPW
Patient simulator
Parts
Order No.
Notes
Recorder paper
74205
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......