B30 Patient Monitor
4-6
Document no
.
2044677-001
2.1 NET section troubleshooting
Problem
Cause
What to do
Monitor does not connect to
the network.
Monitor connects to the
network, but disconnects
unexpectedly (‘Network
down: central serve’
message on the monitor
screen).
Patch panel
Patch cable not connected to HUB or to panel.
Patch cable
Patch cable or connector defective.
HUB not connected to power supply.
HUB port closed due to physical layer problems.
HUB port temporarily closed and reopened due
to physical layer problems.
Hubs not properly connected to each other.
Monitor-Network cable
Cable not properly connected to the wallplate
or to the monitor.
Cable or connector defective.
Network cable (inside the
monitor) defective
Replace the Network cable.
Network Connector Board
defective
Replace the Network Connector Board Unit.
Net section of the CPU board
The NET section is defective. Replace the CPU
board.
NET section memory on the
CPU board
The SRAM of the NET section is defective or
uninitialized. The NET cannot be used. See
network service page for details.
Identification plug
There is no identification plug attached to the
monitor.
The identification plug is defective or
uninitialized. The plug cannot be used.
Multi I/O adapter faulty
Replace multi I/O adapter.
‘Check network connectors’
message shows on the
monitor screen
Monitor-Network cable
Cable not properly connected to the wallplate
or to the monitor.
Cable or connector defective.
Network cable (inside the
monitor) defective
Replace the Network cable.
Network Connector Board
defective
Replace the Network Connector Board Unit.
Multi I/O adapter faulty
Replace multi I/O adapter.
Identification plug
There is no identification plug properly attached
to the monitor.
The identification plug is defective or
uninitialized. The plug cannot be used. See
network service page for details.
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......