Electrical Safety Tests
3-7
Document no. 2044677-001
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(5) Read and record the current leakage indicated on the tester.
(6) Change leakage tester switches to:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(7) Read and record the current leakage indicated on the tester.
(8) Change leakage tester switches to:
•
Polarity - NORMAL
•
Neutral - OPEN
•
GND (Earth) - CLOSED
(9) Read and record the current leakage indicated on the tester.
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.
2.7.1 Acceptance criteria NC (Normal condition)
•
(USA only) 300 µA, and the device under test is powered from 100-120 V/50-60 Hz
•
(USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
•
500 µA, and the device under test is powered from a non-centertapped, 200-240 V/50-60
Hz, single-phase circuit
2.7.2 Acceptance criteria SFC (Single fault condition) - ground (earth), line or
neutral open
•
(USA only) 300 µA, and the device under test is powered from 100-120 V/ 50-60 Hz
•
(USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
•
1000 µA
NOTE: Center-tapped and non-center-tapped supply circuits produce different leakage
currents and the UL and IEC limits are different.
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......