B30 Patient Monitor
5-22
Document no
.
2044677-001
NOTE: Messages ‘Zeroing’ and ‘Zeroed’ is shown in the NIBP message field and next to the
Calibrate
selection momentarily. After this, a pressure bar will appear beside the menu and the
text ‘Calibrating’ is shown in the NIBP digit field.
4.
Calibration:
−
Connect the NIBP hose to the module connector.
−
Connect an external manometer with a pump to both tubes of the hose.
−
Pump about 200 mmHg pressure.
−
Verify that both pressure values, B1 and B2, shown in the prompt field of the
calibration menu match the manometer reading. If not, adjust the by turning the
ComWheel.
−
Press
ComWheel
to complete the calibration.
NOTE: Messages ‘Calibrating” and ‘Calibrated’ are shown in the NIBP message field and next to
the
Calibrate
selection.
NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the
ComWheel, even in cases where the value would be correct. For example, change the value
one step higher and then back one step lower. The ‘Calibrated’ text should appear in the
display. This ensures that the calibration procedure is correctly registered and stored by the
module.
5.
Change the
Protection
setting from OFF to ON to disable
Calibrate
selection - the color of
the
Calibrate
selection changes from white to grey.
6.
Press the NIBP module buttons
Auto ON/OFF
and
Start Cancel
simultaneously for 3
seconds. This disables the
Protection
selection - the color of the
Protection
selection
changes from white to grey.
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......