B30 Patient Monitor
A-4
Document no
.
2044677-001
5.
Click
Open License
button, choose the license file you have. The
CPU Serial Number
and
Software Version
will be automatically imported through the license file.
6.
Press
Emergence
button to start software download.
NOTE: When the emergence download for the main software is done successfully,
waiting for 3 to 5 minutes until the B30 monitor restart automatically.
Verification and configuration
1.
Perform Factory Reset.
Monitor Setup
Install/Service
(16-4-34)
Service
(26-23-8)
Set/Test
Factory Reset
After the Factory Reset is completed, the B30 monitor will restart automatically.
2.
Set the time and date.
Monitor Setup
Time and Date
3.
If necessary, set the monitor’s network communication according to the used network
software.
Monitor Setup
Install/Service
(16-4-34)
Service
(26-23-8)
Frame
Network
Dri Config
DRI level
Central station software S-CNET01 -> DRI Level = 2001
Central station software S-CNET02 -> DRI Level = 2002
Central station software L-NET03 or L-NET05 -> DRI Level = 2003
NOTE: If the DRI level is changed, the B30 monitor will restart automatically. The
change will take effect only after restart.
4.
Verify that the software has been installed correctly.
Monitor Setup
Install/Service
(16-4-34)
Service
(26-23-8)
SW Version / Unit Id
5.
Verify that the active language is installed and activated correctly
Monitor Setup
Install/Service
(16-4-34)
Service
(26-23-8)
SW Settings
Country Settings
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......