Planned maintenance
3-31
Document no. 2044677-001
2.
Service reset
Check the Service Reset. Disconnect the power cord and remove the batteries to reset the
monitor. Switch the monitor back on and check that the monitor performs a Cold Start, i.e.
all trend information is cleared.
NOTE: Cold start have three ways:
−
Disconnect all power source and reset the monitor
−
When the monitor power on, hold on the
Power
hardkey 15 seconds
−
Turn off the monitor, waiting for 15 minutes
"
3.
Watchdog
Enter the
Set/Test
service menu and perform
Watchdog
Check that the monitor restarts.
"
4.
Service Log reset
Enter the
Service Log
service menu.
Clear the content of the Service Log by selecting
Reset Log
from the menu.
"
5.
Electrical safety check
Perform an Electrical safety check and a leakage current test. Check that the monitor and
all connected units function normally after the performed test.
"
6.
Final cleaning
Switch off the monitor and perform final cleaning.
Fill in all necessary documents.
"
Содержание b30
Страница 2: ......
Страница 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Страница 13: ...1 Introduction...
Страница 14: ......
Страница 61: ...2 Installation...
Страница 69: ...3 Maintenance...
Страница 70: ......
Страница 105: ...4 Troubleshooting...
Страница 106: ......
Страница 125: ...5 Service Menu...
Страница 126: ......
Страница 151: ...Parameters 5 25 Document no 2044677 001 4 4 4 NIBP Buttons Leds Buttons Leds Data Not in use in B30 Monitor...
Страница 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Страница 159: ...6 Field replaceable unit...
Страница 160: ......
Страница 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Страница 183: ...Frame disassembly 6 23 Document no 2044677 001...
Страница 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Страница 191: ...7 Technical specification...
Страница 192: ......
Страница 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Страница 199: ...8 N F C REC Module...
Страница 200: ......
Страница 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Страница 244: ...B30 Patient Monitor C 2 Document no 2044677 001 Notes Signature Functional Inspection OK N A Fail...
Страница 251: ......