Function of pressure monitors integrated in
Column valves V9-C or V9H-C
If Column valve C, labeled V9-C or V9H-C, is used, the two pressure monitors integrated in
the valve can be used to monitor the column pressure.
Column valve V9-Cs and V9H-Cs does not contain pressure monitors.
Note:
The pressure sensors integrated in the inlet and outlet ports of Column
valve C measure the pressure before the column, Pre column pressure (A),
and the pressure after the column, Post column pressure (B).
The Delta column pressure (C), which is the pressure drop over the column,
is calculated as the difference between the Pre column pressure and the
Post column pressure.
Pressure alarms can be set for both the Pre column pressure and the Delta
column pressure. Pressure control of the flow may be based on either the
Pre column pressure or the Delta column pressure.
ÄKTA pure User Manual 29119969 AB
69
2 The ÄKTA pure instrument
2.4 Instrument modules
2.4.12 Pressure monitors