System settings - Air sensor
The table below describes the air sensor related system settings available for ÄKTA pure.
Normal (30 µl) is used to detect when a buffer or sample vessel is empty. High (10 µl) is used to detect
small air bubbles
Description
Instruction name
Alarm inlet A air sensor enables or disables the air sensor alarm
for the built-in air sensor at inlet A. If the alarm is enabled and air
is detected, an alarm will be triggered and the method will be
paused.
Alarm inlet A air sensor
Alarm inlet B air sensor enables or disables the air sensor alarm
for the built-in air sensor at inlet B. If the alarm is enabled and air
is detected, an alarm will be triggered and the method will be
paused.
Alarm inlet B air sensor
Alarm inlet S air sensor enables or disables the air sensor alarm
for the built-in air sensor at the sample inlet. If the alarm is enabled
and air is detected, an alarm will be triggered and the method will
be paused.
Alarm inlet S air sensor
Alarm external air sensor X enables or disables the alarm for the
optional air sensor number X. If the alarm is enabled and air is
detected, an alarm will be triggered and the method will be
paused.
Alarm external air sensor X
Sensitivity inlet A air sensor is used together with the Alarm inlet
A air sensor instruction and sets the sensitivity of the built-in air
sensors at inlet A.
Sensitivity inlet A air sensor
Normal (30 µl) is used to detect when a buffer or sample vessel
is empty. High (10 µl) is used to detect small air bubbles
Sensitivity inlet B air sensor is used together with the Alarm inlet
B air sensor instruction and sets the sensitivity of the built-in air
sensors at inlet B.
Sensitivity inlet B air sensor
Normal (30 µl) is used to detect when a buffer or sample vessel
is empty. High (10 µl) is used to detect small air bubbles
Sensitivity inlet S air sensor is used together with the Alarm inlet
S air sensor instruction and sets the sensitivity of the built-in air
sensors at the sample inlet.
Sensitivity inlet S air sensor
Normal (30 µl) is used to detect when a buffer or sample vessel
is empty. High (10 µl) is used to detect small air bubbles
ÄKTA pure User Manual 29119969 AB
455
9 Reference information
9.8 System settings
9.8.5 System settings - Air sensor