7.3.1
Change pump rinsing solution
Maintenance interval
Replace the pump rinsing solution in the system pumps and the sample pump (optional module) every
week to prevent bacterial growth.
Required material
The following material are required:
• 20% ethanol
• Syringe, 25-30 ml
Illustration of the system pump piston rinsing
system
The illustrations below show the parts and tubing of the system pump piston rinsing system
1
2
3
4
5
Description
Part
Rinsing system tube holder, top
1
Rinsing system tube
2
Rinsing system tube holder, bottom
3
Outlet tubing
4
Inlet tubing
5