7.6.10
Clean the Conductivity flow cell
Maintenance interval
Clean the Conductivity flow cell when required.
Required material
The following materials are required:
• Luer connector
• Waste container
• Syringe, 25-30 ml
• 1 M NaOH
• Distilled water
Instruction
Follow the instruction below to clean the flow cell of the Conductivity monitor.
Action
Step
Disconnect the fingertight connector and the piece of tubing from the top of the Conduc-
tivity monitor, and attach a Luer connector.
1
Disconnect the piece of tubing from the bottom of the Conductivity monitor, and connect
a piece of waste tubing to the Conductivity monitor. Place the waste tubing in a waste
container.
2
300
ÄKTA pure User Manual 29119969 AB
7 Maintenance
7.6 Maintenance when required
7.6.10 Clean the Conductivity flow cell