Superloop
Possible cause and action
Problem
• The Superloop is filled to the max. Pressure is not released
anywhere.
Manually turn Injection valve to Manual load position.
Overpressure during filling
Cabinet
Possible cause and action
Problem
• The air intake on the back or on the left side of the instrument
is covered
Make sure that none of the air intakes on the instrument are
covered.
• Broken fans
Contact Service.
• Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
The temperature of the in-
strument or an instrument
component is too high
• The waste tubing from the Buffer tray is loose
Contact Service.
Liquid from the Buffer tray
is flowing onto the table
Power and ICU
Possible cause and action
Problem
• The power cord is not connected
Connect the power cord to the wall outlet and to the electrical
inlet on the instrument. Make sure that the cord is attached using
the clip, thereby preventing the cord from coming loose.
• No electric current in the wall outlet
Make sure that there is electric current in the wall outlet.
• A fuse in the instrument ICU is broken
Contact Service.
• The instrument is overheated
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The instrument cannot be
turned on
396
ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.7 Troubleshooting: Other components