Tubing and pressures
The table below shows the tubing to disconnect when checking and calibrating the pressure monitors.
The UNICORN names of the pressures measured by the monitors are also shown.
Pressure in UNICORN
Tubing to disconnect
Pressure monitor
System pressure
Tubing from the System pump pressure monitor
System pump pressure
monitor
Sample pressure
Tubing from the Sample pump pressure monitor
Sample pump pressure
monitor
PreC pressure
Tubing to Column valve V9-C or V9H-C port In
Pre-column pressure
monitor
306
ÄKTA pure User Manual 29119969 AB
7 Maintenance
7.7 Calibration procedures
7.7.2 Calibrate the pressure monitors