Pump
Action
Description
Error code
Check that there is no blockage of the pump
outlet. Restart the instrument. If recurrent please
contact Service.
Internal pump error
51-53
See Section 8.6 Troubleshooting: Pumps, on
page 387
High temperature
54
UV monitor U9-M
Action
Description
Error code
Restart the instrument. If recurrent please con-
tact Service.
Grating error
21, 25, 26,
31
Restart the instrument. If recurrent please con-
tact Service.
Block filter error
22, 23, 32
Restart the instrument. If recurrent please con-
tact Service.
Internal instrument error
24
Restart the instrument. If recurrent please con-
tact Service.
Spectrum calibration error
27
Restart the instrument. If recurrent please con-
tact Service.
Lamp error
28, 29
See UV monitor U9-M and UV detector unit, on
page 360.
High temperature
30
UV detector U9-D
Action
Description
Error code
Restart the instrument. If recurrent please con-
tact Service.
Internal instrument error
24, 26, 28,
29, 31, 33
Check that the detector and flow cell are fitted
correctly. If warning reappears, contact Service.
Too low light intensity
25
Check that the detector and flow cell are fitted
correctly. If warning reappears, contact Service.
Too high light intensity
27
Check that the detector is fitted correctly. If
warning reappears, contact Service.
Too high light intensity, R channel
30
406
ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.10 Error codes