
Action
Step
Empty and discard the remaining tubes:
4
• Press the QuickRelease device to the lock position, and empty the remaining tubes.
• Pull the QuickRelease device to the release position, and discard the tubes.
Discard the tubes
Empty the tubes
Fraction collector tubes and bottles
The tubes and bottles used in Fraction collector F9-C must fulfill the requirements listed in the table
below. Examples of manufacturers are also listed in the table.
Examples of manufac-
turers
Height (mm)
Diameter (mm)
Tube or bottle
size (ml)
Max.
Min.
Max.
Min.
NUNC™
56
50
11.5
10.5
3
VWR™
76
70
12
10.5
5
BD™ Biosciences, VWR
102
96
13.3
12
8
BD Biosciences
120
114
17
16
15
BD Biosciences
116
110
30
28
50
Nalgene™, Kautex™
121
-
L: 64.5
W: 64
1
L: 55
W: 55
1
250 ml bottle
1
Length and width of the rectangular bottle base
100
ÄKTA pure User Manual 29119969 AB
3 ÄKTA pure external modules
3.2 Fraction collector F9-C
3.2.3 Cassettes, Cassette tray and racks
Содержание AKTApure
Страница 1: ...ÄKTA pure User Manual ...
Страница 2: ...Page intentionally left blank ...
Страница 520: ...Page intentionally left blank ...
Страница 521: ...Page intentionally left blank ...