3.4 Sample pump S9 and S9H
Introduction
This section describes the design and function of Sample pump S9 and S9H.
Function of the Sample pump
The Sample pump is dedicated to direct loading of sample onto a column, or to filling of sample loops
or Superloops. The pump consists of two pump heads that work alternately to give a continuous, low
pulsation, liquid delivery. There are two sample pump configurations available, one for ÄKTA pure 150
and one for ÄKTA pure 25. To ensure delivery of correct liquid volume, the pumps must be free from
air. Each pump head is equipped with a purge valve that is used for this purpose.
The Sample pump also comprises a pressure monitor.
The table below contains the operating limits and labels of the sample pump configurations.
Max. pressure
Flow rate
Pump type
Pump label
Configuration
10 MPa
0.001 to 50 ml/min
P9-S
P9-S
Sample pump S9,
ÄKTA pure 25
5 MPa
0.01 to 150 ml/min
P9H
P9H S
Sample pump S9H,
ÄKTA pure 150
Front view
1
2
3
4
5
6
7
8
9
108