9.8.4
System settings - Pressure alarms
The table below describes the pressure alarm related system settings available for ÄKTA pure.
Description
Instruction name
Alarm system pressure sets the alarm limits for the system
pressure. When enabled and the system pressure falls outside
the set pressure limits, an alarm will be triggered and the method
will be paused. Default values for the alarm limits are set by the
values in the column list when a column is selected in the method
and Alarm system pressure is set as a variable. Low alarm is
only triggered if the pressure first exceeds the Low alarm limit for
ten seconds continuously and then falls below the Low alarm
limit.
Note:
Setting the Low alarm or the system flow rate to 0 deactivates the
low pressure alarm.
Alarm system pressure
Alarm sample pressure sets the alarm limits for the sample
pressure. When enabled and the sample pressure falls outside
the set pressure limits, an alarm will be triggered and the method
will be paused. Default values for the alarm limits are set by the
values in the column list when a column is selected in the method
and Alarm sample pressure is set as a variable. Low alarm is
only triggered if the pressure first exceeds the Low alarm limit for
ten seconds continuously and then falls below the Low alarm
limit.
Note:
Setting the Low alarm or the system flow rate to 0 deactivates the
low pressure alarm.
Alarm sample pressure
Alarm delta column pressure sets the alarm limits for the delta
column pressure (pre-column pressure minus post-column pres-
sure). When enabled and the delta column pressure falls outside
the set pressure limits, an alarm will be triggered and the method
will be paused. Low alarm is only triggered if the pressure first
exceeds the Low alarm limit for ten seconds continuously then
falls below the Low alarm limit.
Note:
• Setting the Low alarm to 0 deactivates the low pressure alarm.
• Instruction Alarm delta column pressure is available only when
Column valve V9-C or V9H-C (5-columns) is selected in the
component list.
Alarm delta column pressure
ÄKTA pure User Manual 29119969 AB
453
9 Reference information
9.8 System settings
9.8.4 System settings - Pressure alarms