Description
Instruction name
Stop fractionation frac 2 ends the fractionation after the set delay
volume for the second Fraction collector (specified in System
Settings:Tubing and Delay volumes) has been collected. The
outlet valve is then turned to position Waste.
Note:
If Stop fractionation frac 2 is issued when both Fractionation
frac 2 and Peak fractionation frac 2 are active, fractionation is
stopped after the set delay volume has been collected. The Outlet
valve V9-O or V9H-O remains in position Outlet 10 / Frac 2 (or
Outlet 1 if Outlet valve V9-Os is mounted on the instrument) and
peak fractionation in the second Fraction collector continues.
Stop fractionation frac 2
Peak fractionation frac 2 enables collection of only those peaks
that fulfill the conditions set in the Peak fractionation parameters
instruction.
Peak fractionation frac 2
Stop peak fractionation frac 2 ends the peak fractionation in
second Fraction collector after the set delay volume (specified in
System Settings:Tubing and Delay volumes) has been collected.
The outlet valve is then turned to position Waste.
Stop peak fractionation frac 2
Sets fraction numbers to restart from 1 for the second Fraction
collector. The restart occurs when the instruction is issued. The
instruction overrides the continuous numbering mode if Fraction-
ation numbering mode frac 2 is set to Continue in System Set-
tings.
Reset frac number frac 2
During fractionation or peak fractionation the instruction Feed
tube frac 2 moves the second Fraction collector tube rack forward
one tube after the set delay volume has been collected and a
fraction mark is set. When fractionation or peak fractionation is
not ongoing, Feed tube frac 2 moves the rack instantly and no
fraction mark is set.
Feed tube frac 2
Applicable if no fraction collector is used. Fractionation in outlet
valve enables fractionation via the outlet valve. When the set
fraction size/outlet has been collected, the outlet valve turns to
the next position. A fraction mark is set in the chromatogram for
each new outlet position. If using Outlet valve V9-Os or V9H-Os
(1-outlet), only one fraction can be collected via Outlet 1 position.
Fractionation in outlet valve
Applicable if no fraction collector is used. Stop frac in outlet valve
ends the fractionation in outlet valve after the set delay volume
(specified in System Settings:Tubing and Delay volumes) has
been collected. The outlet valve is then turned to position Waste.
Stop frac in outlet valve
476
ÄKTA pure User Manual 29119969 AB
9 Reference information
9.9 Manual instructions
9.9.4 Manual instructions - Fraction collection