Fill the Superloop using System pump A
Follow the instruction below to fill the Superloop using the System pump A. Note that Mixer valve has
to be installed for this loop filling technique.
NOTICE
Glass tube splinter. Make sure to set the system pressure below the max pressure
of the Superloop before executing a flow, when the Superloop is connected.
The Superloop can also be filled as part of a method run, as set in the Sample Application
phase in the Method Editor. For multiple injections, it may be more convenient to fill the Super-
loop once, as described in the instruction below.
Tip:
Action
Step
In the Manual instructions dialog:
1
• Select Alarms:Alarm system pressure.
• Set Mode as Enabled.
• Set a High alarm level that is below the maximum pressure of the Superloop.
• Click
.
In the Process Picture:
2
• Click the Injection valve and select Sample pump load.
Result: The Injection valve switches to Sample pump load position.
ÄKTA pure User Manual 29119969 AB
187
5 Operation
5.7 Sample application
5.7.2 Sample application using a Superloop
™