Step
Action
1
Empty the air trap completely from liquid using the Drain instruction for
the air trap (see
).
2
Use the instruction System
→
Calibrate
→
LSL166_air to calibrate the low
level sensor using air.
3
Use the instruction System
→
Calibrate
→
LSH167_air to calibrate the high
level sensor using air.
4
Fill the air trap until liquid comes out through the top valve by pumping in
liquid using the Fill instruction.
5
Turn off the pump.
6
Use the instruction System
→
Calibrate
→
LSL166_liquid to calibrate the
low level sensor using liquid.
7
Use the instruction System
→
Calibrate
→
LSH167_liquid to calibrate the
high level sensor using liquid.
7 Maintenance
ÄKTA ready XL Operating Instructions 29281616 AE
145