Step
Action
5
Remove the screw cap that covers the clamp ring screw (D) using the tool
illustrated.
D
E
6
Take the 8 mm HEX attachment head and fit it to the tool as illustrated be-
low. Hold the pressure plate and while removing the four hex screws hold-
ing the pressure plate in place using the tool. Then remove the pressure
plate (E).
7
Wrap new PTFE tape around the threads of the four hex screws to prevent
thread seizure.
5 Preparation
5.2 Install disposable flow kits
ÄKTA ready XL Operating Instructions 29281616 AE
101