Step
Action
5
For the isocratic setup, connect the air trap tubing (A) to the pump cham-
ber outlet labeled OUT with a TC clamp with the tubing size adapter (B).
A
B
The illustration above shows the isocratic setup.
6
For the gradient setup, connect the air trap tubing to the jumper tubing
without using the tubing size adapter.
5 Preparation
5.2 Install disposable flow kits
ÄKTA ready XL Operating Instructions 29281616 AE