Problems solution
8
Problems solution
This section describes the error signals and provides recommendations for troubleshooting.
Recorder Troubleshooting
error signal
Cause
Recommendation
At
LED event keys
At
LED event keys
blink once per second.
An acoustic signal sounds
error when trying to start
recording.
The voltage disposable battery
or accumulator is too low.
The disposable battery is low or the
battery is not sufficiently charged, or
a stack has been inserted wrong
type. Place a new disposable alkaline
battery ouum sufficiently charged
NiMH battery.
The
LED electrodes
The
LED electrodes
blink in yellow, once per
second. An acoustic signal
error if you try to start
recording.
At least one electrode is loose or
ECG cable is not connected.
Connect the electrodes to the loose ECG
cable, or connect the ECG cable to the
recorder.