Introduction
18
SEER ™ 1000
2067634-131 Revision E
Certification
Medical equipment
Certification with respect to electrical shock, fire and mechanical
hazards only in accordance with IEC 60601-1 CAN / USA C22.2 NO.
601.1, IEC 60601-1-2 and IEC 60601-2-47.
Accuracy of the input signal
The amplitude frequency response corresponds to the following requirements of IEC
60601-2-47 Section 51.5.9:
•
In the burner, the effect of a rectangular pulse 5mV / 100 ms after the pulse does not
lead to a displacement of position zero amplitude greater than 0.1 mV in relation to the
zero line before drive. The reduction of the pulse is less than 0.3 mV / s. The excess of
the first pulse flank is less than 10%.
•
Playback of all pulses of a triangular pulse sequence of 1.5 mV / 40 ms, which simulates
a next series of R-waves is within 80% to 110% reproduction of a sequence of triangular
pulse 1.5 mV / 200 ms.
Warning EMS / CEM / AF
The system was designed and tested to meet the requirements for electromagnetic compatibility
(EMC). System changes or additions to this system not expressly approved by the manufacturer
may cause EMC issues with this or other equipment.
high frequency devices may impair the use or the accuracy of the device or system. During the
installation and use of the device or system should be observed known AF sources in the
environment, including:
•
radio and television stations,
•
portable and mobile devices for wireless communication (cell phones, radios)
and
•
ray machines ray, CT scan or MRI machines