Maintenance
9
Maintenance
Perform the following maintenance procedures as described.
Perform daily by visual inspection. If verified that any parts need to be repaired, contact an
authorized service representative GE Healthcare so that repair work can be carried out.
Attention indications
ATTENTION
EQUIPMENT DAMAGE - Some chemicals can damage the plastic housing of the device
and cables. Moreover, some substances are conductive or electrically insulating, and may
affect the signal quality.
•
Use only detergents and disinfectants according to the manufacturer's instructions.
Follow especially all the rules necessary for dilution.
•
Squeeze out excess disinfectant cloth before use.
•
Never immerse the appliance, or the ECG cable ECG electrode lines in liquids. This can lead
to corrosion of the metal contacts and affect the signal quality.
•
The device connection contacts should not accumulate liquid. In such cases, absorb
the liquid with a soft, lint-free cloth.
•
Do not use, otherwise, ether solvents such as acetone or gasoline.