Problems solution
104
SEER ™ 1000
2067634-131 Revision E
Bluetooth connection failures
error signal
Cause
Recommendation
Error connecting
recorders and applications
The recorder and the application of
the device are very close or very
distant
For best results, the devices do not
place at a distance of 0.5 m between
them at a distance of not more than 10
m between them.
The Bluetooth function of the
application device is not connected.
Make sure that the Bluetooth
function of the device application is
turned on.
Objects between the recorder and the
application of the device damage the
connection.
Remove all line of sight of the objects
between the recorder and an
application device.
AF Interference Device
environmental harm the
connection.
Remove any devices that emit
interference AF vicinity of the
recorder device and application.
After the start of
recording is not active any
wireless connection.
The Bluetooth function of the device
is not connected. It is turned off after
recording starts.
Press and hold the event button for
about 3 seconds, until a series of
beeps indicates that the wireless
connection is enabled.