Maintenance
112
SEER ™ 1000
2067634-131 Revision E
•
To clean the device, the ECG cable or line electrodes should not ever be used conductive
solutions, or solutions containing chlorides, wax or wax components.
•
Do not use solutions or products, containing ammonium chloride compounds, such
as:
O benzyl ammonium chloride Dimethyl
O benzyl ammonium chloride Dimethyl
O Solutions containing quaternary ammonium chloride
O Solutions containing quaternary ammonium chloride
O Solvents or abrasive cleaners of any type
O Solvents or abrasive cleaners of any type
O Acetone
O Acetone
O ketone
O ketone
O betadine
O betadine
O Sodium Salts
O Sodium Salts
•
Device, cable lines and electrodes must never be sterilized by autoclaving or
steam.
•
Do not use any of the accessories, if they are not completely dry.
Clean and disinfect recorder
Note the following for the recorder cleaning:
•
Remove the battery before cleaning the device.
•
Clean the device before disinfecting the surface.
•
For the external cleaning of the device, use a cloth, lint-free, lightly moistened with water and
mild soap solution.
•
Disinfect the device before first use, before passing the device on to another person and at
regular intervals.
Use the following disinfectants in accordance with the APIC guidelines (1996) for the
selection and use of disinfectants:
O Sodium hypochlorite (domestic bleach 5.2%), minimal dilution
O Sodium hypochlorite (domestic bleach 5.2%), minimal dilution
1: 500 (at least 100 ppm free chlorine) and maximum dilution of 1:10.